38
Chapter 5:
Performance Testing
Continuous Wave (CW) Doppler Imaging
Image Quality Verification Test/Livescan
• Products with replaced subassemblies, or products that have been otherwise disassembled, must undergo an
Image Quality Verification Test/Livescan.
• The Image Quality Verification Test/Livescan should be performed after successfully completing all applicable
performance tests listed prior in this chapter.
• The test is completed before returning the system to service.
• A certified sonographer must perform the test.
• The Livescan test performed is at the discretion of the Sonographer and will represent their acceptance of a
successful service event.
• Review all saved images and verify that the images are displayed properly.
Printer
Test CW
Doppler
Imaging
1
Attach the P21x transducer.
2
Press the
Patient
key.
3
Select the
Cardiac
exam type.
4
Press the
Done
softkey.
5
Press the
Doppler
key for the Doppler sample gate.
6
Press the PW softkey to switch to CW Mode.
7
Press the
Doppler
key again for the Doppler spectral trace.
8
Place a large drop of ultrasound gel on the transducer lens.
9
Adjust the
Gain
control as necessary and then gently tap the top of the gel and observe a
reflection on the spectral trace and the sound from the speakers.
10 Press the
Freeze
key and then save the image. Press the
Freeze
key again to return to live
imaging.
11 Press the
2D
key to return to 2D imaging.
Test Printer
Operation
1
Verify proper printer type is configured in the system Setups page.
1
Press the print button and verify that the printer begins to print an image. After the image
begins to emerge from the printer, press the print button again. The printer should ignore
the second print command.
2
Verify the proper content of the printed image.
Содержание M-Turbo
Страница 1: ...M Turbo Ultrasound System Service Manual TM...
Страница 6: ...2 Chapter 1 Introduction...
Страница 20: ...16 Chapter 3 Troubleshooting...
Страница 36: ...32 Chapter 4 Replacement Procedures...
Страница 44: ...40 Chapter 5 Performance Testing...
Страница 48: ...44 Appendix A Replacement Parts List Figure A 1 Power Supply P08850 Figure A 2 Speaker Assembly P03872...
Страница 52: ...48 Appendix B Service Event Report Service Event Report Form...
Страница 56: ...52 Index...
Страница 57: ......