20
Chapter 4:
Replacement Procedures
Display Replacement
Test the Display
7
Remove the four screws from the Display Hinges per
.
Figure 4.6 Display Screws
Screws (4)
Display
Replacement
1
Set the new display in place.
2
Install the four hinge screws that hold the Display in place. Torque the screws to 5.5 inch
pounds.
3
Reinstall the Display Rear Enclosure, screws (2) and screw caps.
4
Connect the two connectors that connect the Display to the Main PCBA.
5
Place the Control Panel in place.
6
Reinstall the four screws that hold the Control Panel in place. Torque the screws to 5.5 inch
pounds.
Test Display
1
Replace the battery or attach an external power supply.
2
Press the
Power
key to apply power to the system.
3
Verify the display operates correctly.