Chapter 4:
Replacement Procedures
31
Main PCBA
Replacement
Replace the Main PCBA by following the reverse of the removal procedure. Do not tighten all the
screws until everything is in place.
1
Replace the Main PCBA.
2
Reinstall the Nest Frame Assembly. The Nest Frame Socket Head Cap Screws should be
torqued to 4.5 inch pounds
3
Reconnect the speaker wires.
4
Reinstall the Power Supply PCBA.
5
Reinstall the SD Card Daughter-card and copper tape.
6
Reinstall the TGC assembly.
7
Reinstall the shield to the Power Supply.
8
Tighten all screws to their specified torque of 5.5 inch pounds.
9
Reinstall the Control Panel.
10 Reinstall the bottom enclosure.