Chapter 4:
Replacement Procedures
17
Chapter 4: Replacement Procedures
Display Replacement
Required Parts
Service Assembly, LCD Display, M-Turbo (P08659)
Required Tools
• #1 Phillips screwdriver
• Torque screwdriver, 2.0–10.0 inch pounds (0.23–1.1 newton meter)
• An anti-static mat
• A wrist grounding strap
Display Removal
Caution:
Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first.
Initial ESD symptoms may be slightly degraded performance or image quality.
Caution:
All fasteners should be torqued to 5.5 inch pounds except where noted.
Display
Removal
1
Remove the battery from the system.
2
Remove the two screws from the back of the system per
Figure 4.1 System Rear
Screws (2)