Chapter 7: Replacement Procedures 47
2. Secure the four screws on the replacement Display Subassembly hinges to the
top housing with a #1 Phillips screwdriver (Figure 7.3). Torque screws to
6.1 in./lb. (0.7 newton meter).
3. Place the Control Panel Subassembly onto the Display Subassembly.
4. Install the display wire harness and flex circuits.
Note: Check that the reference designators on the flex circuits match the reference
designators on the PCBA connectors.
5. Replace the 100-pin flex circuit on the bottom module (Figure 7.5).
Figure 7.5
Placing the Control Panel Subassembly onto the Display
6. Carefully replace the Control Panel Subassembly by inserting the tabs on the top
of the Control Panel Subassembly into the slots.
7. Set the Control Panel into place and replace the two screws inside the battery
compartment (Figure 7.6). Torque screws to 7.1 in./lb. (0.8 newton meter).
8. Place the battery in the battery compartment.
9. Turn on the system.
10. Perform the Display tests in
verify that the Display Subassembly is functioning properly.
Control Panel
Trackball
Display Wire
Harness
Anti-static mat
Display
100-pin flex
circuit
Содержание C1.9 PLUS
Страница 1: ...C1 9 PLUS Ultrasound System Service Manual...
Страница 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Страница 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Страница 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Страница 98: ...90 C1 9 PLUS Ultrasound System Service Manual...