62
C1.9 PLUS Ultrasound System Service Manual
5. Carefully mate the bottom housing to the top housing and secure the eight
screws using a #1 Phillips screwdriver (Figure 7.23). Torque the screws to
7.1 in./lb. (0.8 newton meter).
Figure 7.23
Securing the Bottom Housing to the Top Housing
6. Place the battery into the battery compartment.
7. Turn on the system.
8. Verify that the Main PCBA Subassembly functions properly by performing the
tests in
.
7.5
Transducers
The transducers have no user-serviceable parts; return transducers
to SonoSite for exchange.
7.6
AC Power Adapter
The AC power adapter has no user-serviceable parts. Return power adapters to
SonoSite for exchange.
Screws (8)