Chapter 8: Performance Tests 67
8.4.2
Axial Measurement Accuracy
Note: Measurements must be performed while the image is frozen.
To test axial accuracy:
1.
Acquire the image.
2.
Press
Freeze
.
3.
Press
Measure
. Two calipers appear on the image display. A menu appears, on
which are listed two
distance
icons, an
ellipse
icon, a
delete
icon, and a
calcs
icon, if applicable. (If the
caliper line
setup is on, then a dotted line connects the
two calipers. See the
SonoSite Ultrasound System User Guide
, if necessary.) The
first caliper in the menu is active by default.
4.
Use the
trackball
to position one of the calipers.
5.
Press
Select
to fix the caliper and enable the other caliper.
6.
Use the
trackball
to move the other caliper. The results update as you move the
caliper, and the measurement is complete when you finish moving the calipers.
(Press
Select
to alternate the active caliper, and adjust the measurement with the
trackball
.)
7.
To perform another distance measurement on the image, select the other
distance
icon and repeat the preceding steps.
8.
Measure the distance, center to center, of two pins that are 5-12 cm apart
vertically.
9.
Verify that the distance measured is within the tolerance listed in Table 8.1.
8.4.3
Lateral Measurement Accuracy
To test the lateral measurement accuracy:
1.
2.
Measure the distance, center to center, of two pins that are 4-10 cm. apart
horizontally.
3.
Verify that the distance measured is within the tolerance listed in Table 8.1.
4.
Press
Freeze
to return the system to live 2D mode.
Table 8.1
System Measurement Accuracy
MEASUREMENTS
TOLERANCE
Axial Distance
+/- 2%
Lateral Distance
+/- 2%
Содержание C1.9 PLUS
Страница 1: ...C1 9 PLUS Ultrasound System Service Manual...
Страница 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Страница 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Страница 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Страница 98: ...90 C1 9 PLUS Ultrasound System Service Manual...