Chapter 7: Replacement Procedures 49
7.2.3
Removing the Control Panel Subassembly
1. Press the battery release on the lower right side of the system to remove the
system battery.
2. Remove the two screws in the battery compartment with a #1 Phillips
screwdriver; this releases the Control Panel Subassembly from the top housing
(Figure 7.6).
Figure 7.6
Removing the Control Panel Subassembly
3. Insert an anti-static mat between the Control Panel Subassembly and the Display
Subassembly to prevent damaging the Display LCD.
4. Carefully lay the Control Panel Subassembly over onto the display to expose the
display wire harness and the two 40-pin flex circuits.
5. Carefully disconnect the 100-pin flex circuit from the bottom module.
6. Disconnect the display wire harness and the two 40-pin flex circuits from the
Control Panel Subassembly (Figure 7.7).
7. Remove the Control Panel Subassembly.
Screws (2)
Battery
compartment
Содержание C1.9 PLUS
Страница 1: ...C1 9 PLUS Ultrasound System Service Manual...
Страница 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Страница 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Страница 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Страница 98: ...90 C1 9 PLUS Ultrasound System Service Manual...