PAT-806-IT
– USER MANUAL
63
unloaded output voltage: 4 V ... 12 V AC
Test
current: ≥ 200 mA for R = 0.2 Ω ... 1.99 Ω
adjustable upper limit in the range of: 10
mΩ …1.99 Ω with resolution 0.01 Ω
adjustable measuring time: 1 s
…60 s with a resolution of 1 s, and "Cont" mode (continu-
ous measurement) at point 0
Measurement resistance of protective conductor I=10 A (only Protection Class I)
Display range
Resolution
Basic uncertainty
0 m
Ω...999 mΩ
1
mΩ
(3 % m.v. + 4 digits)
(3 % m.v. + 40 digits)*
1.00
Ω ... 1.99 Ω
0.01
Ω
* for two-wire measurement
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0.1
%/ºC
unloaded output voltage: <12 V AC
Test current: ≥ 10 A for R ≤ 0.5 Ω
adjustable upper limit in the range of: 10
mΩ …1.99 Ω with resolution 0.01 Ω
adjustable measuring time in the range of: 1 s
…60 s with resolution of 1 s
Measurement resistance of protective conductor I=25 A (only Protection Class I)
Display range
Resolution
Basic uncertainty
0 m
Ω...999 mΩ
1
mΩ
(3 % m.v. + 4 digits)
(3 % m.v. + 40 digits)*
1.00
Ω ... 1.99 Ω
0.01 Ω
* for two-wire measurement
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0.1
%/ºC
unloaded output voltage: <12 V AC
test current: ≥ 25 A for R ≤ 0.2 Ω
adjustable upper limit in the range of: 10
mΩ …1.99 Ω with resolution 0.01 Ω
adjustable measuring time in the range of: 1 s
…60 s with resolution of 1 s
Measurement of insulation resistance using test voltage of 100 V
Test range according to IEC 61557-2 for U
N
= 100 V: 100
kΩ…99.9 MΩ
Display range
Resolution
Basic uncertainty
0 k
Ω...1999 kΩ
1
kΩ
(5 % m.v. + 8 digits)
2.00 M
Ω ... 19.99 MΩ
0.01
MΩ
20.0 M
Ω ... 99.9 MΩ
0.1
MΩ