4.3 Compensation of the test lead resistance during the measure-
ment of the protective lead resistance using 200 mA current (au-
to-zero)
In
R
PE
200mA
mode, press
. Auto-zero
screen will appear with blinking symbol
.
Use the probe connected to
I2
socket and touch PE pin of
the test socket.
Press
START
, to begin auto-zero.
After finishing auto-zero process, for 1s is
displayed message
and the meter enters
the measuring function.
Completed auto -zero process is signalled during the measurement by displaying
.
In order to remove auto-zero proceed in the same manner, but disconnect the test lead from PE.
After finishing, for 1s is displayed message
.