Section
AirWave System Operator’s Manual
48
NOTICE! Please follow hospital guidelines and procedures for suctioning patients when suctioning the
AirWave Sensor.
4. ETT Movement Smoothing
In the event that there is artifact in the echo waveform from the patient airway, the ETT movement indicator may
display “jumpy” results. In extreme cases, this may result in nuisance alarms for high or low tube tip movement.
The AirWave provides a feature to smooth out these jumpy readings.
To enable this smoothing filter, navigate to the
Main Menu
,
and select “
ETT Movement Smoothing
”.
Figure I-2
Change the setting to “On” on the screen that appears (see Figure I-2). Exit and accept the changes. Both the
Home and Waveform screens will now display “SM On” to indicate the feature is enabled (see Figure I-3). On the
Home screen, the indicator is located near the upper left corner of the screen; on the Waveform screen, the
indicator is located above the airway waveform.
Figure I-3
To disable this smoothing filter, navigate to the
Main Menu
, and select “
ETT Movement Smoothing
”.
Change the setting to “Off” on the screen that appears. Exit and accept the changes.
NOTICE! Once enabled, the ETT Movement Smoothing filter will remain in effect until turned off by the
user.
NOTICE! When the ETT Movement Smoothing filter is enabled, there will be a time delay between gross
movements of the ETT and when the AirWave shows the movement.