Desaturation [% SpO
2
]
The minimum decrease in SpO
2
for a desaturation event to be detected.
4)
Max. Drop (Artefact) [% SpO
2
per s]
An artefact will be detected if desaturation decreases faster than this predefined value.
5)
Oscillation Range [% SpO
2
]
A new desaturation will be scored if this predefined value is exceeded.
Fig. 6.16: SpO2 Analysis Oscillation Range (OR)
6)
Min. SpO
2
(Artefact) [%]
If SpO
2
is smaller than this defined value, an artefact is detected.
7)
Body Position Correlation [s]
A desaturation will be defined separately and will be defined as a desaturation caused by a Body
Position change, whenever it occurs after a body position change within the predefined time frame.
6.2.3.9 CPAP Analysis
Fig. 6.17: CPAP Analysis
The trend of the CPAP pressure is calculated in 15-second intervals and displayed.
“Steps displayed in report” allows to choose the raster for pressure values displayed in the report.
6.2.4 Menu - Analysis Channels
Within the Menu
Analysis Channels
the presentation of the analysis window can be formatted.
The Channel Order can be changed by using the Drag & Drop function.