6.2.3.2 Classification of Arousals
Fig. 6.6: Classification of arousals
Arousals are labelled in the selected source channel. As soon as an arousal is detected, the analysis
“Arousal Classification”
will check if an event occurred in any of the other channels within the delay
time set here. The delay time can be adjusted as required and can be set separately for each of the
following events: Flow, Desaturation and HR-Acc/Decc. Depending on their priority the arousals will be
correlated to these events. Using the drag & drop function, the order within the list can be changed
and by these means the priority of classification is altered.
“Fixed Duration [s]”
In the
“Keys”
tab you can set the duration of an arousal for quicker manual editing. Define a key for
the event
“Arousal fixed duration”
.
Fig. 6.7: Key for arousal duration
6.2.3.3 Activity Analysis
Fig. 6.8: Activity Analysis
Time-frame [s]
Size of the Analysis-Timeframe in detection of Activity.
Max. value: Activity is equal to the maximal value within the timeframe.
Average value: Activity is equal to the value composed of the average value within the timeframe.