Part No. 502621 Issue 10
25
7.0: Specifications (continued)
7.0: Specifications (continued)
7.4: Notes
1. Primary voltage absolute value and drift is not
specified. The OD4 uses ratiometric techniques
to compensate for primary voltage drift.
2. The way in which the OD4 functions means
a special configuration must be used for
transducers outside of the standard range.
This is selectable by links. The majority of
Solartron LVDT transducers are within the
standard range. See section 5.
3.
OD4 can drive into a 1 kΩ load but this offers
no advantage. 10-100 kΩ is recommended.
4. Output voltage range can be adjusted as
required anywhere within this range by using
a combination of gain and offset, for example.
±10 VDC, ±5 VDC, 0-5 VDC, 0-10 VDC,
4-20 mA.
5. Current output may be used at the same time
as voltage output. Calibration of voltage and
current cannot be individually adjusted.
6. Protection applies to the product when fully
installed according to the user manual. During
installation the top of the enclosure has to be
removed for access to user adjustments. At this
time standard ESD handling precautions for
electronic equipment should be followed.
7. The OD4 complies with the toughest electrical
emissions and immunity regulations. Compliance
requires installation according to the user manual.
Compliance does not guarantee performance as
the installation environment may be outside of test
specification limits. The flexibility of OD4 means it
can be installed in a variety of ways according to
user requirements. Simple installations with short
non-screened cables will meet the lesser light-
industrial immunity regulations. Heavy industrial
installations, especially with longer cables, will
need more careful installation with screened
cables.