21
6.0: Half-Bridge only Variant
6.0: Half-Bridge only Variant
Part No. 502621 Issue 10
This is a half-bridge optimised variant of the
standard product. The excitation frequency is
higher (see specification) and the transducer
input connector is wired to accept half-bridge
transducers with standard connections.
6.1: Connecting the transducer
The 5-pin DIN plug is screwed into the case
mounted socket. Transducers not fitted with a
plug should be wired to the plug supplied. LVDT
transducers cannot be connected to this input.
6.2: Setting up the Half-Bridge Transducer
The setting up procedure is the same as LVDT
transducers. See section 5. The sensitivity of
half-bridge transducers is generally lower than
for LVDT types, the x2 and x4 gain position may
have to be used. A list of standard link settings
for all Solartron Transducers is available from the
downloads section of www.solartronmetrology.com.
Red
Blue
Yellow
+
-
Case
Half-Bridge
Cable Screen
Black
Case
Plug view
into pins
Standard Half-Bridge G auging Probe Plug Connections
Note 1: + indicates inward m ovem ent of the tip.
Note 2: The transducer body m ay be disconnected from the cable screen by
cutting the black wire inside the connector
4
2
1
3
5
LINK
CO
RE
Blue
Yellow
R ed
V
V
A
B
Centre Tap
V
O
Half-Bridge Electrical Connections
Red and Blue
Energising
Yellow Signal
Red and Yellow In Phase for Inward Displacement
Black
Transducer Body Ground