Part No. 502621 Issue 10
16
5.0: Setting up the Transducer
and OD4 (continued)
5.0: Setting up the Transducer and OD4 (continued)
5.2: Basic Procedure
To set-up the OD4, some basic steps should
be followed.
The following steps describe a typical setting
procedure and applies to most applications.
Other procedures may be used as appropriate.
For a bi-polar output i.e. ±10 VDC or ±20 mA,
follow steps 1 to 3. For a uni-polar output i.e.
0-10 V, 0-20 mA or 4-20 mA, follow steps 1 to 4.
In either case, step 5 (final checks) should be
followed on the completion of the set-up.
Zero
-5V
+5V
Null
electronics
transducer
Null
Zero
transducer
electronics
Null
+5V
0V
+10V
electronics
transducer
Shift
zero
V/V
±V
Hz
K
Ω
Step 1
Set-up OD4
links
Step 2
Align OD4 and
Transducer Null
Step 3
Set OD4 and
Transducer Range
Step
5
Final Checks
Step 4
Add Offset
if Required