SNIJDERS SCIENTIFIC
B.V.
ISO9001 certified by Lloyd’s
Economic Premium ECP01E, March ‘06
Page 17 of 22
Subject to change
6.4.
Starting the program with time input
A program can be started at a specific point of time:
-
call up “program selection” with
=> Start program
, followed by
-
select a “program” using the cursor keys
-
push
, followed by
-
-
change to menu “program start”
-
enter start time/ start date or start delay, start segment and remaining segment time
-
start program with
.
6.5.
Alarmfunctions
-
Alarms will occur when the deviation is +/- 5°C from the temperature setpoint and +/- 5% from
the humidity setpoint. For the light and CO
2
injection no alarms can occur.
-
When the internal temperature and humidity deviates from the settemperature and -humidity, the
buzzer, alarm led and the potential free contact (backside cabinet) will be energized.
-
The buzzer is mutable by using the switch on page 7, number 14. The alarm itself stays
energized.
-
When the internal temperature is too low the cooling will be switched off. When the internal
temperature will be too high, the heating, lighting & fan will be switched off.
-
The safety-thermostats, located at the back of the control panel, are set on the absolute minimum
and maximum safety temperature in order to prevent that an alarm is given when switching on
the
unit. Standard these safety-thermostats are set on -5°C & +60°C.