SNIJDERS SCIENTIFIC
B.V.
ISO9001 certified by Lloyd’s
Economic Premium ECP01E, March ‘06
Page 13 of 22
Subject to change
4
Cy = -1 (constantly
repeating program).
Note: below the “CtrlCon” are mentioned 8 digits. The last 4 can be used to switch the
lighting (1 = ON; 0 = OFF).
When you like to use a constantly repeating program (Cy=-1), you have to program the last
segment as follows:
-
set the temperature and/or light level at the required end value
-
set Cy (repeating program) at -1.
50 programs up to 99 segments can be programmed; a total of 1000 segments can be implemented.
Programs are created by programming set points and segments times, segment after segment.
Furthermore, the states of the control contacts 1-3 and the active parameter set, can be defined for
each segment. The setpoint profiles can be output either as a ramp or as a step (configurable).
Output as a ramp has been chosen for the following diagrams.