• Prior to starting infusion, inspect the fluid path for kinks, a
closed clamp, or other upstream occlusions, and remove all
air bubbles to prevent air embolism.
• Close the fluid path tubing with the clamp before removing
the cassette from the pump to prevent unregulated gravity
infusion.
• If you are using a CADD® administration set or CADD
™
medication cassette reservoir that does not have the flow
stop feature (catalog number does not start with 21-73xx):
You must use a CADD® extension set with an integral Anti-
Siphon Valve or a CADD® administration set with either
an integral or an Add On Anti-Siphon Valve to protect
against unregulated gravity infusion that can result from an
improperly attached cassette.
• For detailed instructions and warnings pertaining to
the CADD
™
medication cassette
reservoir or CADD®
administration set, please refer to the instructions supplied
with those products.
• Attach the cassette (the part of the CADD
™
medication
cassette
reservoir or CADD® administration set that attaches
to the pump) properly. An improperly attached or detached
cassette could result in unregulated gravity infusion of
medication from the fluid container or a reflux of blood.
• Do not prime the fluid path with the tubing connected
to your catheter. This could result in over-delivery of
medication or air embolism.
• Ensure that the entire fluid path is free of all air bubbles
before connecting to your catheter to prevent air embolism.
• If the pump is dropped or hit, inspect the pump for damage.
Do not use a pump that is damaged or is not functioning
properly. Contact your clinician for further instructions.