Failure to properly follow warnings, cautions, and instructions
could result in damage to the pump or death or serious injury.
Warnings
• If the pump is used to deliver life-sustaining medication, an
additional pump must be available.
• Use of a syringe with the CADD® administration set may
result in UNDER-DELIVERY of medication. Syringe function
can be adversely affected by variations in plunger dimension
and lubricity, which can result in greater force required
to move the syringe plunger. A syringe plunger will lose
lubrication as it ages and, as a result, the amount of under-
delivery will increase which could on occasion, be significant.
You must regularly compare the volume remaining in the
syringe to the pump’s displayed values such as Reservoir
Volume or Given to determine if under-delivery is occurring
and, if necessary, contact your clinician.
• Do not use rechargeable NiCd or nickel metal hydride (NiMH)
batteries. Do not use carbon zinc (“heavy duty”) batteries.
They do not provide sufficient power for the pump to
operate properly.
• Always have new batteries available for replacement. If
power is lost, non-delivery of medication will occur.
• If the pump is dropped or hit, the battery door tabs may
break. Do not use the pump if the battery door or tabs
are damaged because the batteries will not be properly
secured; this may result in loss of power and non-delivery of
medication. Contact your clinician for further instructions.
• If a gap is present anywhere between the battery door and
the pump housing, the door is not properly latched. If the
battery door becomes detached or loose, the batteries will
not be properly secured; this could result in loss of power
and non-delivery of medication.