Messages and Alarms
When you see and hear:
Take this action:
[No message]
T
wo
-
Tone
alarm
With no AC adapter attached,
the batteries have been
removed while the pump is
running. The pump is now
stopped and unpowered.
Install batteries to silence the
alarm.
OR
Batteries were removed
within approximately
15 seconds after stopping the
pump. Install new batteries, if
desired, to silence the alarm.
Otherwise, the alarm will stop
within a short period of time.
T
wo
beeps
(
long
-
shorT
)
[Screen displays
current pump status]
The disposable (CADD®
administration set or
CADD
™
medication cassette
reservoir) is not aligned
with the pump, or is
damaged, or a malfunction
of the pump sensor(s) is
occurring. Reposition the
pump to silence the alarm. If
repositioning the pump does
not silence the alarm within
two minutes, the pump
will display “No Disposable,
Clamp Tubing.”