Messages and Alarms
When you see and hear:
Take this action:
T
wo
-
Tone
alarm
High Pressure
There may be a kink in the
tubing or a clamp may be
closed. Unkink the tubing
or open the clamp and the
pump will resume delivery.
You may press
⁄
or
„
to stop the pump and silence
the alarm for 2 minutes.
After you remove the cause
of high pressure, start the
pump if necessary. If the
alarm continues, contact your
clinician.
T
wo
-
Tone
alarm
Key pressed,
Please release
If a key is being pressed,
stop pressing it. If the alarm
persists, close the tubing
clamp and remove the pump
from use. Contact your
clinician.
T
hree
Two
-
Tone
beeps
every
5
minuTes
LowBat
The AA batteries are low
but the pump is operable.
Change the AA batteries
soon.
T
wo
-
Tone
alarm
Motor Locked,
remove all power
Batteries are depleted and
the pump was powered up
with the AC Adapter. Install
new AA batteries, reconnect
the AC adapter, and restart
the pump.