This guide is intended to supplement 

instructions on how to operate the pump supplied 

by your Clinician. Perform only those procedures for 

which you have received training by your Clinician.

Legacy

Patient Information

CADD-Legacy

®

 1

Ambulatory Infusion Pump

Model 6400

CONTINUOUS DELIVERY

10011896-001.indd   1

2/27/2018   7:08:09 PM

Содержание 6400

Страница 1: ...lement instructions on how to operate the pump supplied by your Clinician Perform only those procedures for which you have received training by your Clinician Legacy Patient Information CADD Legacy 1 Ambulatory Infusion Pump Model 6400 CONTINUOUS DELIVERY ...

Страница 2: ...talling New Batteries 7 Main Screen 10 Starting the Pump 11 Stopping the Pump 12 Turning the Pump On Off 12 Removing a Cassette 13 Attaching a Cassette 14 Resetting the Reservoir Volume 16 Priming the Tubing Using the Pump 17 Inserting the Tubing into the Air Detector 18 What if I drop or hit the pump 20 Messages and Alarms 21 ...

Страница 3: ...re are other CADD Legacy pump models available review the rear label of the pump to ensure it is a CADD Legacy 1 Model 6400 pump before use This pump delivers medication at a continuous rate in milliliters per 24 hours mL 24hrs The pump stores programmed and historical information that your clinician needs for your specific therapy Your clinician will instruct you on the proper use of this pump Th...

Страница 4: ...as Reservoir Volume or Given to determine if under delivery is occurring and if necessary contact your clinician Do not use rechargeable NiCd or nickel metal hydride NiMH batteries Do not use carbon zinc heavy duty batteries They do not provide sufficient power for the pump to operate properly Always have new batteries available for replacement If power is lost non delivery of medication will occu...

Страница 5: ...ached cassette For detailed instructions and warnings pertaining to the CADD medication cassettereservoir or CADD administration set please refer to the instructions supplied with those products Attach the cassette the part of the CADD medication cassettereservoir or CADD administration set that attaches to the pump properly An improperly attached or detached cassette could result in unregulated g...

Страница 6: ...eries installed Frozen medications must be thawed at room temperature only Do not heat the CADD medication cassette reservoir in a microwave oven as this may damage the medication the CADD medication cassette reservoir or cause leakage Do not immerse the pump in cleaning fluid or water or allow solution to soak into the pump accumulate on the keypad or enter the battery compartment Do not clean th...

Страница 7: ... or CADD administration set that attaches to the pump Display Power Jack Accessory Jack AC Indicator Light Air Detector Keypad CADD Legacy 1 Pump Diagram Display Shows information and messages AC Indicator Light The green indicator light is on when you are using the AC Adapter to power the pump ...

Страница 8: ...air bubbles from the fluid path Œ Used by the clinician Advances you from one screen to the next and silences alarms Allows you to increase a value or scroll up through items on a menu Î Allows you to decrease a value or scroll down through items on a menu Å Allows you to put the pump into a low power state when not in use or back into full power ...

Страница 9: ... do not provide sufficient power for the pump to operate properly which could result in death or serious injury Always have new batteries available for replacement If power is lost non delivery of medication will occur and depending on the type of medication being administered could result in death or serious injury If the pump is dropped or hit the battery door or tabs may break Do not use the pu...

Страница 10: ...tery compartment Insert the batteries over the top of the battery strap The pump will beep if the batteries are inserted correctly 6 Replace the battery door The pump will power up automatically NOTES If you put the batteries in backwards the display will remain blank Reinsert the batteries making sure to match the and markings The power up sequence will start the pump will go through an electroni...

Страница 11: ... door becomes detached or loose the batteries will not be properly secured this could result in loss of power non delivery of medication and depending on the type of medication being administered death or serious injury CAUTION Do not store the pump for prolonged periods of time with the batteries installed Battery leakage could damage the pump ...

Страница 12: ...ou will see on the pump s display most of the time It is called the Main Screen and shows the following RUN ResVol 50 0 ml When running Mode of pump Reservoir Volume When stopped Mode of pump Battery Status for example LowBat Status of Reservoir Volume STOPPED ...

Страница 13: ...d remove all air bubbles to prevent air embolism An undetected upstream occlusion may result in under or non delivery of medication and depending on the type of medication being administered could result in death or serious injury Air embolism can result in death or serious injury 1 Press and hold 2 When you see 3 sets of dashes disappear release After automatically reviewing the program RUN and t...

Страница 14: ...he pump is stopped you may put the pump into a low power state by turning it off The pump may be turned off when it is disconnected from your access site and it is going to be stored To turn the pump off 1 Press and hold Å 2 When you see 3 sets of dots appear release Å To turn the pump on Press and hold Å The pump will power up and automatically review all screens Stopping STOPPED Stopping Turning...

Страница 15: ...se all tubing clamps and disconnect the tubing from your access device as instructed by your clinician 4 Insert a coin into the slot in the lock and turn it clockwise The lock will pop out when you unlock the cassette 5 A continuous alarm will sound and the pump will display No Disposable Clamp Tubing The alarm may be silenced by pressing or 6 Remove the cassette from the pump hinge pins 7 Discard...

Страница 16: ... in death or serious injury For detailed instructions and warnings pertaining to the CADD medication cassette reservoir or CADD administration set please refer to the instructions for use provided with those products CAUTION Frozen medication must be thawed at room temperature only Do not heat the CADD medication cassette reservoir in a microwave oven as this may damage the medication the CADD med...

Страница 17: ...mp and you feel the button click into place WARNING Attach the cassette the part of the CADD medication cassettereservoir or CADD administration set that attaches to the pump properly An improperly attached or detached cassette could result in unregulated gravity infusion of medication from the fluid container or a reflux of blood which could result in death or serious injury 5 Gently twist push a...

Страница 18: ...ress to reset the volume to the programmed volume NOTE When the pump calculates that 5 mL remain in the fluid container beeps sound and ResVol Low appears on the main screen This alarm recurs at every subsequent decrease of 1 mL until the Reservoir Volume reaches 0 mL at which point the pump stops and the Reservoir Volume empty alarm sounds Reservoir Volume 100 0 ml ...

Страница 19: ...After you see Prime and 3 sets of dashes release 4 Press and hold again to fill the fluid path and to eliminate air bubbles The screen displays Priming and you will hear a short beep each time the pump goes through a delivery cycle The pump will stop automatically after priming 10 delivery cycles NOTE The air detector alarm is automatically disabled when priming 5 If the tubing is not yet fully pr...

Страница 20: ...tector is designed to detect air bubbles in the fluid path When the tubing is routed through an Air Detector that is turned on an air bubble exceeding the specified size will cause an alarm to sound and the pump to stop 1 If your clinician instructed you to use the Air Detector make a small loop of tubing underneath the air detector and hold it with your thumb 2 Place the tubing over the groove in...

Страница 21: ...d flat in the groove 4 Follow your clinician s instructions for connecting the tubing to your access site Open all clamps WARNING Ensure that the entire fluid path is free of all air bubbles before connecting to your catheter to prevent air embolism Air embolism could result in death or serious injury Retention nubs ...

Страница 22: ...up with the arrow on the side of the pump Gently twist push and pull on the cassette to make sure it is still firmly attached Check the battery door to make sure it is still firmly attached If the cassette or the battery door is loose or damaged do not use the pump Immediately stop the pump close the tubing clamp and contact your clinician WARNING If the pump is dropped or hit inspect the pump for...

Страница 23: ...rm Then follow your clinician s instructions for removing air Two tone alarm Battery Depleted The AA batteries are depleted Install two new AA batteries New batteries must be installed even when an AC Adapter is connected Two tone alarm Battery Removed Pump won t run When using an AC Adapter the AA batteries have been removed while the pump is running or you have tried to start the pump with deple...

Страница 24: ...e pump if necessary If the alarm continues contact your clinician Two tone alarm Key pressed Please release If a key is being pressed stop pressing it If the alarm persists close the tubing clamp and remove the pump from use Contact your clinician Three two tone beeps every 5 minutes LowBat The AA batteries are low but the pump is operable Change the AA batteries soon Two tone alarm Motor Locked r...

Страница 25: ...ter stopping the pump Install new batteries if desired to silence the alarm Otherwise the alarm will stop within a short period of time Two beeps long short Screen displays current pump status The disposable CADD administration set or CADD medication cassette reservoir is not aligned with the pump or is damaged or a malfunction of the pump sensor s is occurring Reposition the pump to silence the a...

Страница 26: ...A disposable must be properly attached in order for the pump to run Press or to silence the alarm If the alarm persists contact your clinician Two tone alarm No Disposable Pump won t run You have tried to start the pump without a disposable CADD administration set or CADD medication cassette reservoir attached or the disposable is attached but is not aligned with the pump or is damaged or a malfun...

Страница 27: ...Reservoir Volume Empty The Reservoir Volume has reached 0 0 mL Press or to silence the alarm Then install a new fluid container if appropriate and reset the reservoir volume Three single beeps RUN ResVol Low The Reservoir Volume is low Change the fluid container soon See page 16 for further details Two tone alarm Service Due See manual The pump is scheduled for service Press or to silence the alar...

Страница 28: ...s not flowing from the fluid container to the pump Check for a kink in the tubing or a closed clamp between fluid container and pump Press or to stop the pump and silence the alarm for 2 minutes After you remove the cause of upstream occlusion start the pump if necessary If the alarm continues contact your clinician ...

Страница 29: ... Pumpplacementduringbathing showering ______________ _____________________________________________________ Pump placement during sleep _____________________________________________________ _____________________________________________________ Storageofmedication _________________________________ _____________________________________________________ ...

Страница 30: ...10011896 001 28 Notes ...

Страница 31: ......

Страница 32: ...mentioned are the trade names trademarks or service marks of their respective owners 2005 2010 2016 2018 Smiths Medical All rights reserved 2018 02 10011896 001 x Smiths Medical ASD Inc 6000 Nathan Lane North Minneapolis MN 55442 USA Tel 1 800 258 5361 USA CA Tel 1 614 210 7300 European Representative Smiths Medical International Ltd 1500 Eureka Park Lower Pemberton Ashford Kent TN25 4BF UK Tel 44...

Отзывы: