5
Demand Dose
0 to 9.9 ml*
Delivery rate (Continuous Rate + Demand
Dose) programmable from 40 to 125 ml/hr.
Default: 0 ml (See Table 2, 3 & 4 for
Scroll ranges)
*The Maximum Demand Dose is 20 with
software revision E or higher.
Demand Dose Lockout
5 minutes to 24 hours in the following
increments:
• 1 minute for values between
1 and 20 minutes
• 5 minutes between 20 minutes
and 24 hours
Default: 5 min
Set Delivery Limit
0.5 ml to 1000 ml (or the mg or mcg
equivalent), or “No Limit”:
0.01 from 0.01 to 0.1
0.1 from 0.1 to 100
1.0 from 100 to 1000
10.0 from 1000 to 10000
100.0 from 10000 to 100000
1000.0 from 100000 and up
Default: 0.5 ml or mcg or mg equivalent
Given
0 to 99999.99 in 0.01 unit increments.
Clinician Bolus
0.1 ml to 20.00 ml (or mg or mcg equivalent)
Delivery rate (Continuous Rate + Clinician
Bolus): 125 ml/hr nominal (See tables 2, 3 & 4
for Scroll ranges)
High Pressure Alarm
18 ± 9 psi [1.24 ± 0.62 bar]
Air Detector Alarm
Single bubble greater than 0.100 ml
Options Specifications
Lock Level
LL0, LL1, LL2. Default: LL2
Epidural Mode
On or Off. Default: Off
Units*
Milliliters (ml), milligrams (mg),
micrograms (mcg).
Default: milligrams
Time
0000 to 2359
Air Detector
Turned On or Turned Off. Default: On
Biomed Toolbox Specifications
Custom Concentrations
All individual mg or mcg concentration
settings may be enabled or disabled (at least
one concentration must be enabled).
Default: All On
Program Limits*
Maximum program limits may be
programmed for Demand Dose,
Continuous Rate, and Clinician Bolus.
Default: maximum program limits.
Dosing Limit*
Delivery Limit, a Maximum Doses per Hour, or
neither. Default: Neither
Key Beeps
On or Off. Default: On
Res Vol Trip Point
1 to 999 ml in increments of 1 ml, or
“Standard.” Default: Standard
Res Vol Empty Alarm*
Single or Insistent alarm. Default: Single
Pump Stopped Alarm*
Beep or Two‑tone alarm. Default: Beep
AutoLock
Not In Use, LL1 Key/Code, LL2 Key/Code, LL1
No Key or LL2 No Key. Default: Not In Use
PM
(Preventive Maintenance) Reminder
1 to 24 months in 1 month increments, or
“Not In Use.” Default: Not In Use
Custom Lock Level Code
001 to 899 (excluding preset code) in
increments of 1. Default: 061
Units Selection*
Program units to appear in the Programming
Units screens. Default: All Programming Units
Units Location
Options, Program or Biomed Toolbox.
Default: Programming screens
* If programmed to be part of Options settings in the Biomed Toolbox.