22
Remote Dose Cord Check
• Check the Remote Dose Cord button
operation by programming the pump with
the following values:
Units
Milligrams
Concentration
1.0 mg/ml
Continuous Rate
0.0 mg/hr
Demand Dose
1.0 mg
Demand Dose Lockout 0 hrs 5 min
Delivery Limit
No Limit
Reservoir Volume
10.0 ml
Dose Counters
0/0 (Press the
ENTER key to clear
in the Reports
menu)
Milligrams Given
0.00 mg (Press the
ENTER key to clear
in the Reports
menu)
• Press the STOP/START key. Press the YES
key. The pump should sequentially display
all of the programmed values.
• After RUNNING appears on the display,
press the Remote Dose Cord button and
note the time. The pump should beep twice
and begin to deliver. Count the number of
pump activations. The pump should make
ten double activations. After ten double
activations, the display should show a
reservoir volume of 9.0 ml. Press the Remote
Dose Cord button two more times within the
next 5 minutes. The pump should not deliver
and the message “Dose Not Delivered Dose
Locked Out” should be displayed.
Doses Given and Doses Attempted Check
• Stop the pump by pressing the STOP/START
key, then the YES key. Press the REPORTS
key to advance to the Dose Counters report.
The display should show 1/3. (If the above
steps have not been followed exactly,
different values may appear.)
• Press the ENTER key. The display should
now show 0/0.
MG GIVEN Report Check
• Press the REPORTS key to advance to the
Milligrams Given report. The display should
now show 1.0 mg if the Remote Dose cord
was tested. (If the above steps have not been
followed exactly, different values may appear.)
• Press the ENTER key. The display should
now show 0.0 mg.
Air Detector Test (if applicable)
This test will verify the function of the optional
air detector. To perform this test, the
CADD‑Prizm
®
PCS II pump must have an air
detector installed and the air detector must
be turned on. The previous program from
the DOSE key check can be used to perform
this test.
• Attach an empty CADD
™
medication
cassette reservoir or CADD
®
administration
set to the pump.
• Latch and lock the set to the pump.
• Open the air detector door and thread the
tubing through the groove.
• Close the door making sure the tubing does
not get pinched or kinked.
• Start the pump.
• The pump should respond with a continuous
two‑tone alarm and the display should read
Air in line detected, Pump will not run VIEW
to silence,
• Press VIEW/SILENCE to silence the alarm,
and remove the CADD
™
medication cassette
reservoir or CADD
®
administration set.
• Now attach a CADD
™
medication cassette
reservoir containing fluid and a primed
extension set with anti‑siphon valve, or
a primed CADD
®
administration set with
anti‑siphon valve to the pump. Make certain
there is no air in the fluid path.
• Latch and lock the set to the pump.
• Open the air detector door and thread the
tubing through the groove.
• Close the door making sure the tubing does
not get punched or kinked.
• Start the pump.
• Deliver a demand dose. (NOTE five minutes
must have passed since the delivery of the
last demand dose.)
• The pump should deliver the dose without
an air detection alarm.
Occlusion Accuracy Tests
This manual provides two testing options for
the Occlusion Pressure Range Test and the
Accuracy Test. Only one option needs to be
performed. It is not necessary to perform
both Option 1 and Option 2.