4
Specifications (Nominal)
General Pump Specifications
Resolution
CADD
™
medication cassette reservoir
or CADD
®
administration set, 0.050 ml
per pump stroke nominal
Size
4.4 cm x 10.4 cm x 14.1 cm [1.7 in x 4.1 in
x 5.6 in] excluding cassette or other accessories
Weight
568 g [20 oz.] including 9 volt battery and empty
100 ml CADD
™
medication cassette reservoir,
excluding other accessories
Pump Alarms
Low battery power; depleted battery power;
external power source low, faulty, depleted;
pump stopped; pump fault; low reservoir
volume; high delivery pressure; air in line; Air
Detector faulty or detached (only with the use
of the optional Air Detector); Air Detector Port
Cover detached; delivery too slow; key stuck;
cassette detached or unlocked; print failure,
epidural cassette not used.
Bolus Volume at Occlu sion Alarm
Pressure 0.050 ml resolution
administration sets/CADD
™
medication
cassette Reservoirs <0.25 ml
Power Sources
9 volt alkaline or lithium battery such as
DURA CELL
®
Alkaline MN 1604 or ULTRALIFE
®
Lithium U9VL; CADD
®
External Power Source
(EPS) Power Pack reorder number 21‑3801; AC
Adapter. The expected life of a 9 volt battery is 12
hours at 100 ml/hour, or approximately 5 days at
10 ml/day (nominal). This estimate is based on
laboratory tests conducted at room temperature
using a new battery. Actual battery life will
vary depending on the brand of battery, shelf
life, temperature conditions, delivery rate, and
frequency of screen display, backlighting
and printing. It is recommended that a new
9 volt battery be kept available for replacement
if necessary.
An internal battery powers the clock. When
it is depleted, it cannot reliably maintain the
clock time. This battery must be replaced by
the manufacturer. The internal battery has an
expected life of 5 years.
* If programmed to be part of pump programming screens in Biomed Toolbox.
System Operating Temperature
+2°C to 40°C (36°F to 104°F)
System Storage Temperature
‑20°C to 60°C (‑4°F to 140°F)
Power Pack Charging Temperature
+10°C to 35°C (50°F to 95°F)
System DeliveryAccuracy
± 6% (nominal)
System Definition
System is defined as a CADD‑Prizm
®
PCS II
pump with an attached CADD
™
medication
cassette reservoir and CADD
®
extension set with
integral anti‑siphon valve, or an attached CADD
®
administration set with integral or add‑on
anti‑siphon valve.
Delivery Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in 1 ml
increments, displayed in 0.1 ml increments.
Default: 1 ml
Units*
Milliliters (ml), milligrams (mg),
micrograms (mcg).
Default: milligrams
Concentration
Mg/ml:
0.1 to 0.5 mg/ml in increments of 0.1
1 to 15 mg/ml in increments of 1 mg/ml
20 to 100 mg/ml in increments of 5 mg/ml
Default: 100 mg/ml
Mcg/ml:
1 to 15 mcg/ml in increments of 1 mcg/ml
15 to 95 mcg/ml in increments of 5 mcg/ml
100 to 500 mcg/ml in increments of 100
mcg/ml. Default: 500 mcg/ml
Continuous Rate
0 to 30 ml/hr (or the mg or mcg equivalent).
Default: 0 mg/hr (See Table 1 for Scroll ranges)