Error indication function
Error Name
LED display
Type
Troubleshooting
Flow rate upper
limit is exceeded
POWER indicator: Green is ON
FLOW indicator: Red is ON
The applied flow rate is above
approx. 110% of maximum rated
flow rate.
Adjust flow to within the
rated flow range.
Outside the
temperature
measurement
range
POWER indicator: Flashing Red
The fluid temperature is lower
than -10
o
C or exceeds 110
o
C.
Adjust the fluid
temperature to within the
rated temperature range.
"Flow rate upper
limit exceeded"
and "Outside the
temperature
measurement
range" occur
together.
POWER indicator: Flashing Red
FLOW indicator: Red is ON
See above.
See above.
System error
POWER indicator: Red is ON
FLOW indicator: Red is ON
Internal data error has occurred.
Turn the power off and
turn it on again. If the
failure cannot be solved,
contact SMC for repair.
POWER indicator: Red is ON
FLOW indicator: Flashing Red
POWER indicator: Red is ON
FLOW indicator: OFF
The temperature sensor is
damaged.
If the error cannot be reset after the above measures are taken, then please contact SMC
Содержание PF3W5 Series
Страница 12: ...11 No PF OMN0010 J Model Indication and How to Order Metal piping...
Страница 13: ...12 No PF OMN0010 J Polyvinyl chloride piping...
Страница 33: ...32 No PF OMN0010 J Characteristics graph Measurable range of ethylene glycol aqueous solution Reference value...
Страница 35: ...34 No PF OMN0010 J PF3W511 Vinyl chloride piping PF3W521 Vinyl chloride piping...
Страница 38: ...37 No PF OMN0010 J PF3W521 Metal attachment...
Страница 40: ...39 No PF OMN0010 J Flow characteristics of the flow rate adjustment valve PF3W504S PF3W520S PF3W540S...
Страница 48: ...47 No PF OMN0010 J Dimensions of lead wire with M8 connector ZS 40 A...