HRX-OM-W069
Chapter 2 Contact input/output communication /Analog output communication
HRL Series
2.4 Contact output signal
2-13
Table 2.4-4
List of alarm selection
Alarm No.
Indication
Explanation
AL01
CH1 Low Level FLT
CH1 abnormal low tank fluid level
AL02
CH1 Low Level WRN
CH1 low tank fluid level
AL03
CH2 Low Level FLT
CH2 abnormal low tank fluid level
AL04
CH2 Low Level WRN
CH2 low tank fluid level
AL06
Fan Inverter
Fan failure
AL09
CH1 High Temp. FLT
CH1 abnormal rise of circulating
fluid temperature
AL10
CH1 High Temp.
CH1 circulating fluid temperature rise
AL11
CH1 Low Temp.
CH1 circulating fluid temperature drop
AL12
CH1 TEMP READY ALARM
CH1 TEMP READY alarm
AL13
CH2 High Temp. FLT
CH2 abnormal rise in circulating
fluid temperature
AL14
CH2 High Temp.
CH2 circulating fluid temperature rise
AL15
CH2 Low Temp.
CH2 circulating fluid temperature drop
AL16
CH2 TEMP READY ALARM
CH2 TEMP READY alarm
AL17
CH1 HX In High Temp. FLT
CH1 abnormal rise in heat exchanger
inlet temperature
AL18
CH1 Press. Sensor
CH1 failure of circulating fluid
discharge pressure sensor
AL19
CH1 High Press.
CH1 circulating fluid discharge pressure rise
AL20
CH1 Low Press.
CH1 circulating fluid discharge pressure drop
AL21
CH2 Press. Sensor
CH2 failure of circulating fluid
discharge pressure sensor
AL22
CH2 High Press. Error
CH2 abnormal rise in circulating fluid
discharge pressure
AL23
CH2 High Press.
CH2 circulating fluid discharge pressure rise
AL24
CH2 Low Press.
CH2 circulating fluid discharge pressure drop
AL25
CH2 Low Press. Error
CH2 abnormal drop in circulating
fluid discharge pressure
AL26
CH2 Flow Sensor
CH2 failure of circulating fluid discharge
flow sensor
AL27
CH2 High Electric Conductivity
CH2 electric conductivity increase
AL30
Digital Input 1
Contact input 1 signal detection
AL31
Digital Input 2
Contact input 2 signal detection
AL34
Communication
Communication error
AL35
Ambient Temp.
Outside of the ambient temperature range
AL36
Maintenance
Maintenance alarm
AL37
Refrigeration Circuit
Compressor circuit failure
AL38
Sensor
Sensor failure
AL39
Controller
Controller failure
AL40
Compressor Inverter
Compressor inverter error
AL41
Compressor Inverter Comm.
Compressor inverter communication error
AL42
CH1 Pump Inverter
CH1 pump inverter error
AL43
CH1 Pump Inverter Comm.
CH1 pump inverter communication error
AL44
CH2 Pump Inverter
CH2 pump inverter error
AL45
CH2 Pump Inverter Comm.
CH2 pump inverter communication error
Refer to Operation Manual
“Installation
・
Operation
”.
Содержание HRL100-A
Страница 10: ...HRX OM W069 Chapter 1 Read before using 1 4 Touch panel flow HRL Series 1 6...