6 SETUP AND INSTALLATION
Ne
xt, connect the measurement heads to the corresponding adapters at the front side of the
PICO
SCALE
(
fi
gure 5.2, No. 4).
NOTICE
Glass
fi
ber cables are very fragile and sensitive to bending. Please always respect
a minimum bending radius of 30 mm, unless otherwise speci
fi
ed.
NOTICE
Please note, that dirt attached to the
fi
ber end-face can signi
fi
cantly lower the
system
’
s performance or even destroy the
fi
ber connector.
I
n this case, the
PICO
SCALE
and/or the corresponding sensor head has to be sent back to Smar-
Act to get repaired.
To ensure optimal signal quality and also to prevent damage of the
fi
ber surfaces, it is very impor-
tant to clean all connectors
fi
rst. This can be done using the One-Click cleaner pen which is part
of our shipment, cf.
fi
gure 6.1. Open the plastic cap and push the plastic tube onto the glass
fi
ber
end piece until you hear a click. To clean the connectors on the
PICO
SCALE
Controller or vacuum
feedthroughs, remove the plastic tube on top of the
fi
ber optic cleaner and then repeat the same
process.
Figure 6.1: (left) Closeups of a clean (a) and dirty (b)
fi
ber endface. Cleaning of the connectors and
sockets using the One-Click pen is illustrated on the right hand side.
Figure 6.1 (a) shows a close-up of a clean end-face, while image (b) shows dirt accumulated on
the end-face. On the right hand side cleaning of the
fi
ber using the One-Click pen is illustrated.
For a detailed description of the One-Click pen please refer to the brief application note delivered
together with the tool.
When the
fi
ber surfaces have been cleaned, gently plug the
fi
ber connectors into the correspond-
ing sockets. The key of the connectors must slide into the matching slot in the socket (
fi
g. 6.2).
Then fasten the connectors by turning the coupling ring.
44
PicoScale User Manual