LD301 – Operation and Maintenance Instruction Manual
6.2
Zero Adjustment Limits
Calibrated span shall not be less than 0.0085 URL and shall not exceed 2 URL.
Low range value shall not be below LRL.
Upper range value shall not be greater than URL.
(LRL = -URL for all models, except absolute, where LRL
vacuum).
Temperature Limits
Ambient :
-40
to
85
°
C
(-40 to 185
°
F)
Process :
-40
to 100
°
C (-40
to 212
°
F)
(Silicone Oil).
0
to
85
°
C (32
to 185
°
F)
(Fluorolube Oil).
-40
to 150
°
C (-40
to 302
°
F)
for LD301L.
-25
to
85
°
C (-13
to 185
°
F)
(Viton O-Rings).
Storage :
40
to
00
°
C (-40
to 212
°
F)
Digital Display :
-10
to
60
°
C (14
to 140
°
F)
-40
to
85
°
C (-40
to 185
°
F)
without damage.
Failure Alarm
In case of sensor or circuit failure, the self-diagnostics drives the output to 3.6 or 21.0 mA,
according to the user's choice.
Turn-on Time
Performs within specifications in less than 5.0 seconds after power is applied to the transmitter.
Volumetric Displacement
Less than 0.15 cm
3
(0.01 in
3
).
Overpressure and Static Pressure Limits
From 3.45 kPa abs. (0.5 psia)* to:
8 MPa (1150 psi) for range 1.
16 MPa (2300 psi) for ranges 2, 3 & 4.
32 MPa (4600 psi) for models H & A5.
40 MPa (5800 psi) for model M5.
52 MPa (7500 psi) for model M6.
* except the LD301A model.
Flange Test Pressure: 60 MPa (8570 psi).
For ANSI/DIN Level flanges (LD301L models):
CLASS
RANGE
TEMPERATURE
150 lb
6 psia to 275 psi (-0,6 to 19 bar)
38ºC
300 lb
6 psia to 720 psi (-0,6 to 50 bar)
38ºC
PN10/16
-60 kPa to 1,4 Mpa
120ºC
PN25/40
-60 kPa to 4 MPa
120ºC
These overpressures will not damage the transmitter, but a new calibration may be necessary.
Humidity Limits
0 to 100% RH.
Damping Adjustment
Through the programmer: adjustable for any value greater than or equal to 0 seconds,
added to the sensor response time (0.2 seconds)
Configuration
Can be done through digital communication using the Hart Protocol or, partially, through local
adjustment.
0.9915
NEW
Содержание ld301
Страница 14: ...LD301 Operation and Maintenance Instruction Manual 1 8...
Страница 18: ...LD301 manual de Instru es Opera o e manuten o 2 4 Fig 2 3 LD301 Software Block Diagram...
Страница 50: ...LD301 Operation and Maintenance Instruction Manual 4 16...
Страница 55: ...Maintenance Procedure 5 5 Fig 5 1 Exploded View...
Страница 62: ...LD301 Operation and Maintenance Instruction Manual 5 12...
Страница 70: ...APPENDIX A...