LD301 – Operation and Maintenance Instruction Manual
5.2
ERROR MESSAGES
POTENTIAL SOURCE OF PROBLEM
SV OUT OF LIMITS
•
Temperature out of operating limits.
•
Temperature sensor damaged.
PV OUT OF LIMITS
•
Pressure out of operation limits.
•
Sensor damaged or sensor module not connected.
•
Transmitter with false configuration.
LOWER RANGE VALUE TOO HIGH
•
Lower value exceeds 24% of the Upper Range Limit.
LOWER RANGE VALUE TOO LOW
•
Lower value exceeds 24% of the Lower Range Limit.
UPPER RANGE VALUE TOO HIGH
•
Upper value exceeds 24% of the Upper Range Limit.
UPPER RANGE VALUE TOO LOW
•
Upper value exceeds 24% of the Lower Range Limit.
UPPER & LOWER RANGE VALUES
OUT OF LIMITS
•
Lower and Upper Values are out of the sensor range limits.
SPAN TOO SMALL
•
The difference, between the Lower and Upper values is less than the 0.75 x (minimum span).
APPLIED PRESURE TOO HIGH
•
The pressure applied was above the 24% upper range limit.
APPLIED PRESURE TOO LOW
•
The pressure applied was below the 24% lower range limit.
EXCESS CORRECTION
•
The trim value entered exceeded the factory-characterized value by more than 10%.
PASSED PARAMETER TOO LARGE
•
Parameter above operating limits.
PASSED PARAMETER TOO SMALL
•
Parameter below operating limits.
Table 5.1 - Error Messages and Potential Source
DIAGNOSTIC WITH THE TRANSMITTER
Symptom
:
NO LINE CURRENT
Probable Source of Trouble
:
Transmitter Connections
•
Check wiring polarity and continuity.
•
Check for shorts or ground loops.
•
Check if the power supply connector is connected to main board.
Power Supply
•
Check power supply output. The voltage must be between 12 and 45 Vdc at transmitter
terminals.
Electronic Circuit Failure
•
Check the main board for defect by using a spare one.
Symptom
:
NO COMMUNICATION
Probable Source of Trouble
:
Terminal Connections
•
Check the terminal
interface connection of the
configurator.
•
Check if the interface is connected to the wires leading to the transmitter or to the terminals
[
+
] and [
-
].
•
Check if the interface is models IF3 (for Hart protocol).
Transmitter Connections
•
Check if connections are according to wiring diagram.
•
Check if there is resistance in the 250
Ω
line.
Power Supply
•
Check output of power supply. The voltage at the LD301 terminals must be between 12 and 45
Vdc, and ripple less than 500 mV.
Содержание ld301
Страница 14: ...LD301 Operation and Maintenance Instruction Manual 1 8...
Страница 18: ...LD301 manual de Instru es Opera o e manuten o 2 4 Fig 2 3 LD301 Software Block Diagram...
Страница 50: ...LD301 Operation and Maintenance Instruction Manual 4 16...
Страница 55: ...Maintenance Procedure 5 5 Fig 5 1 Exploded View...
Страница 62: ...LD301 Operation and Maintenance Instruction Manual 5 12...
Страница 70: ...APPENDIX A...