SkyVision Linx 300 Pre-Install • REV2
SECTION 3. INSTALLATION RESPONSIBILITIES
3-1. SKYTRON Responsibilities
1. SKYTRON requires a minimum of 60 days delivery time from receipt of signed drawings and
OPI document for manufacturing of the proposed SkyVision system.
2. SKYTRON will provide customer with the appropriate signal cable and wiring for the system
described in this proposal.
3. SKYTRON will provide customer with a suggested conduit layout drawing for the system described
in the proposal.
NOTICE
An inspection and evaluation of the installation site must be performed and
the SkyVision Pre-Installation Evaluation form must be completed prior to
the arrival of installation technicians (see attached sample form).
4.
SKYTRON will provide the appropriate number of specialized technical staff to complete the
cable terminations, equipment rack assembly, and testing of the completed SkyVision system.
5. Installation per unit includes two (2) full days of clear access to each room for cable pulling and
terminations, plus at least two (2) days for fitting of connectors, final system configuring, system power
up, and complete system testing.
6.
Installation should take place on consecutive days and must be coordinated by a customer
representative.
7. Installation will include mounting the SkyVision rack if necessary.
3-2. Customer Responsibilities
1. Customer will provide SKYTRON with the appropriate room drawings to include plan, electrical,
and mechanical drawings.
2. Customer will be responsible for supply and installation of all conduit and electrical junction
boxes.
3. Customer will provide proper HVAC ventilation of NDC compatible with system requirement.
4. Customer will comply with grounding requirements.
5. Customer will provide a dedicated and isolated electrical supply to the SkyVision system.
6.
Customer will be responsible for supplying all specified junction boxes.
7. Customer will be responsible for supply and installation of drywall and backing for installation of
wall or ceiling equipment respectively.
8. Customer will be responsible for supply and installation of appropriate nurse documentation
station(s).
9.
Customer will provide SKYTRON with all customer provided equipment source signal specifications
including signal output type, power requirements, and cable connector types.
10.
Customer will provide SKYTRON with required IP addressing information as specified on the
OPI document prior to production.
11. Customer will provide SKYTRON with the dates for cable installation, boom and light installation,
and agree on specific dates for SkyVision system installation. Notices for cable installation and system
installation must be given no less than 21 days prior to start of work.
Содержание Linx 300
Страница 1: ...Pre Installation Manual SkyVision Linx 300 Integration Control System TEC R 0049 REV2 08 05 2015 ...
Страница 4: ...Page 2 SkyVision Linx 300 Pre Install REV2 ...
Страница 26: ...Page 24 SkyVision Linx 300 Pre Install REV2 NOTES ...
Страница 27: ...Page 25 SkyVision Linx 300 Pre Install REV2 NOTES ...
Страница 28: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906 ...