Page 11
SkyVision Linx 300 Pre-Install• REV2
b. Optional Remote Trigger Image/Video Capture
Capture devices specific to endoscopic cameras allow the surgeon to capture images or video using the
trigger on the camera. This method is limited to capturing only the camera source as the surgeon would
need to have the camera in hand when capturing. This is not a standard option to integrate with SkyVision
because of the many endoscopic vendors. Additionally, the SkyVision foot pedal option (SVIS-FTSW or
SVIS-WFTSW) can be used to capture any source that is set as the "Capture Source" at the touch panel.
The foot pedal is an available option in a wired or wireless model. If capturing endoscope images with
the endoscope trigger is required, this must be specified by the sales representative as a "special" on the
OPI documentation. The hospital must supply SKYTRON with the make and model number of the camera
and capture device so SKYTRON can verify compatibility, proper operation, and provide the required
items to quote.
c. Optional Keyboard/Mouse Switch
The KM Switch (SVIS-KM) allows a single keyboard and mouse to control up to four (4) PCs or one DCSS
and three (3) PCs.
When a DCSS is purchased in addition to the on-board PC, the KM Switch must be quoted. If the hospital
uses their PCs instead of purchasing the SkyVision PCs, they may use their keyboard and mouse instead
of the SkyVision keyboard and mouse. The hospital computer can be placed anywhere in the room with
their keyboard, mouse, and monitor. If they would like to use the SkyVision keyboard and mouse, a KM
Switch is required and the hospital PC must be located within 10 feet of the rack.
d. Additional Touch Panels
Optional second and/or third touch panels (SVIS-SECTP) may be selected with a SkyVision system
to enable the system to be operated from more than one nurse documentation station location. In this
configuration, the second and/or third touch panels are routed through the primary touch panel through
VNC connections.
2-7. Remote Rack Locations
The stock cable length for a SKYTRON SkyVision System installation is 100 feet. 200 and 300 feet cable
lenghs may be selected from the OPI document if longer cable lengths are required.
For installations requiring greater distances than 300 feet, consult with the SkyVision Product Manager
so the options can be specified and project details/requirements resolved.
Содержание Linx 300
Страница 1: ...Pre Installation Manual SkyVision Linx 300 Integration Control System TEC R 0049 REV2 08 05 2015 ...
Страница 4: ...Page 2 SkyVision Linx 300 Pre Install REV2 ...
Страница 26: ...Page 24 SkyVision Linx 300 Pre Install REV2 NOTES ...
Страница 27: ...Page 25 SkyVision Linx 300 Pre Install REV2 NOTES ...
Страница 28: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906 ...