QuickTake 15 Performance Profile
Flow range:
5 to 15 L/min
Flow compensation
accuracy:
± 5% of set flow
Compensating flow
back pressure range:
15 L/min at 8 inches water back pressure
10 L/min at 15 inches water back pressure
5 L/min at 20 inches water back pressure
Power:
Rechargeable 7.2-volt lithium-ion (Li-Ion) battery (
see Li-Ion Battery
Shipment on page 4
), or battery with standard AC charger/adapter
(115 V), or battery with deluxe AC charger/adapter (100-240 V)
Battery recharge time:
Approximately 5 hours
Typical Run time:
VersaTrap Cassette: 4 hours at 15 L/min
Results when tested with a new pump and new fully charged
battery. Pump performance may vary.
The AC charger/adapter can be used to extend battery run
time, but it is not a battery eliminator. Therefore, it will
not provide indefinite run times.
Temperature:
Operating:
32 - 113 F (0 - 45 C)
Storage:
-4 - 113 F (-20 - 45 C)
Charging:
32 - 113 F (0 - 45 C)
Protect sample pump from weather when in use outdoors.
Run time features:
User-selectable features, user-adjustable presets (
see Advanced
Setup
)
Preset timed runs:
1, 2, 5, 10, or 15 min
Manually set continuous
run with manual shut-off:
1 to 999 min (repeats 1 to 999-min
runs indefinitely until user stops
pump or power supply is depleted)
Intermittent sampling:
See Advanced Setup
Flow fault features:
If the pump is unable to compensate due to excessive back pressure,
the pump will go into flow fault mode
(see page 12 for details)
. Flow
fault features are enabled as the default setting. Flow fault may be
disabled at the user’s option.
See Advanced Setup.
Flow fault:
Flow fault displays immediately.
Pump shuts off after 5 seconds
Fault restart:
Pump attempts restart every
10 seconds up to five times
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (
inches water
)
Flow Rate (L/min)
5.0
8.0
10.0
12.0
15.0
Filter/Pore Size (µm)
37-mm MCE/0.8
11
18
22
28
36
37-mm PVC/5.0
4
7
9
11
15