• To navigate presets and displays:
Press
or
to scroll through presets or displays.
• To select a run time preset and run the pump:
Scroll to desired run time preset.
Select preset by pressing [
].
The pump will start to run.
• To run the pump:
Press [
].
• To place the pump in hold:
Press [
].
• To repeat a sample run:
From Done, reset (rSEt), or sampling error (SErr), press [
] to return to
run time preset. Press [
] to sample.
•
To turn off LED:
Pump automatically turns off LED after it is idle for 4 minutes. User can
reactivate the LED by pressing and holding
.
•
To interrupt run, terminate a sample, or reset pump:
From Hold, press
. Sample reset (rSEt) will display. Press [
] to reset
pump or
to ignore reset and return to Hold.
• To set flow or calibrate:
See Calibration and Setting Flow for details.
1. Connect a flowmeter to the pump inlet.
2. Scroll to any run time preset. Press and hold CAL for 2 seconds to enter
calibration mode (pump will start running).
3. Press
or
to change flow rate until desired flow appears on the
calibrator (
Flow rate does not display on LED
).
4. Press
and Stor will display. Press [
] to save setting or
to ignore
changes.
Selecting the
Available Presets
Press
or
Press
Press
15
10
5
2
1
Intt
Cont