skcinc.com
Form 37171 Rev 200630
Page 32
5. Gently insert the belt clip screw through the opening in the belt clip and into the pump case.
Tighten the screw until engaged. Do not tighten completely.
6. Replace the screen cover.
a. Place the screen cover and mounting block so that the two holes are aligned with the
holes in the top of the belt clip. Insert the two screws into the mounting block and tighten
until snug.
b. Ensure that the screen cover closes properly.
7. Tighten the screw under the belt clip until snug.
Replacing the Inlet Housing and/or Inlet Filter
1. Remove the four screws from the inlet housing.
1. Pull the inlet housing away from the pump.
2. Remove the O-ring and filter.
3. Insert the new or existing filter and O-ring into the inlet recess. Ensure that the O-ring is fully flat.
4. Align the new or existing inlet housing with the inlet recess.
5. Insert the four screws into the inlet housing. Tighten the screws only until the gap between the
inlet housing and pump is closed.
TROUBLESHOOTING
If the pump is not responding to touch or the pump screen displays uncommon characters, remove and
reinstall the battery (
see Replacing the Battery Pack
). If these problems persist, contact SKC.
Pump Service
Pumps under warranty should be sent to SKC Inc. for servicing. See Limited Warranty and Return Policy.
User may replace external components such as the inlet filter, battery, screen protector,
and/or belt clip. Service must be performed by SKC to maintain performance and intrinsic
safety rating. Warranty is void if pumping compartment is opened by user.