4 Operation
Operating Instructions TENEO
SIVISION digital video system
Intermittent operation of the intraoral camera
4.11.2.2
Functional description
The camera generates digital image data with the help of a CCD sensor. The
camera allows for the acquisition of intraoral and extraoral images.
4.11.2.3
Connecting the SiroCam digital intraoral camera
Connecting the intraoral camera
The intraoral camera is connected on the bottom side of the dentist element.
¾
Plug the connector of the intraoral camera into the correspondingly
marked sockets on the dentist element.
ª
The plug locks in place.
Removing the intraoral camera
The plug of the intraoral camera is secured against falling out.
¾
Grasp the plug by its locking devices
A
and pull it downward.
CAUTION: The intraoral camera is designed for intermittent
operation in dental offices.
In continuous operation, the intraoral camera may become too hot for patients
or users.
¾
To prevent high temperatures, the camera must be deposited in its
holder for a period of t2=3 min following an operating time of t1=1 min.
A
Lens window