5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Sanitize the treatment center
Operating Instructions TENEO
9
Sanitation phase 1 is highlighted on the touchscreen.
9
The
Insert instruments in sanitation bowl
display appears on the
touchscreen.
1.
Pull the tumbler holder
(A)
off of the water unit.
2.
Place the sanitation bowl with cover onto the cuspidor bowl. Make sure
that the sanitation bowl is not contaminated.
3.
Plug the water guide
(B)
of the tumbler filler onto the cover of the
sanitation bowl.
4.
Insert the adapter for the SPRAYVIT valve body
(C)
, the instrument
couplings
(D)
and the suction hoses
(E)
in the sanitation bowl as far as
they will go.
5.
Insert the adapter for the saliva ejector hose
(F)
in the adapter of the
suction hoses
(E)
.
6.
Remove the SPRAYVIT sleeves from the valve bodies, the straight and
contra-angle handpieces from the sanitizable instruments, and the
suction cannulae from the suction hoses.
E
D
C
B
A
F
NOTE: Arrangement of adapters
The adapters for the instrument couplings are color coded:
white = turbine
green = SL and BL motors
blue = BL ISO motor (ISO 3964 interface)
red = SIROSONIC TL scaler
To prevent the hoses from twisting and stretching, the arrangement of the
adapters from left to right must correspond to the arrangement of the
instruments on the dentist element.