11 Troubleshooting
Sirona Dental Systems GmbH
Operating Instructions PerioScan
61 29 626 D 3496
32
D 3496.201.02.05.02
11
Troubleshooting
Display
Error description
Action to be taken by the dentist
Service
Immediately after power-on –
Switched back on too quickly
Wait for at least 30 s between switch-off and
switch-on
Service
Immediately after switch-on – Board connecting
cable defective or replace board pair
Inform service engineer
Start screen
appears until
released
Foot switch actuated during power-on
Switch on again without actuating foot switch
Flashing tip
Worn tip (SIROPERIO 4 PS or SIROPERIO PE 8)
Loose tip
No tip
Wrong tip
Broken tip
The torque wrench is still attached to the
SIROPERIO 4 PS or SIROPERIO PE 8 tip
Handpiece supply cable not connected correctly
Check length with tip gauge, replace tip if neces-
sary
Tighten tip using the torque wrench
Screw tip on and tighten it using the torque wrench.
As soon as any tip other than the tip of the
SIROPERIO 4 PS or SIROPERIO PE 8 is used,
the detection function must be deactivated (see
page 21)
Screw new tip on and tighten it using the torque
wrench
Pull off the torque wrench
Connect handpiece supply once again
Содержание PerioScan
Страница 2: ......
Страница 4: ......
Страница 49: ......
Страница 54: ...Contents Sirona Dental Systems GmbH Operating Instructions PerioScan 61 29 626 D 3496 4 D 3496 201 02 05 02 ...
Страница 95: ......
Страница 100: ...Table des matières Sirona Dental Systems GmbH Notice d utilisation PerioScan 61 29 626 D 3496 4 D 3496 201 02 05 03 ...
Страница 141: ......
Страница 146: ...Índice Sirona Dental Systems GmbH Instrucciones de uso PerioScan 61 29 626 D 3496 4 D 3496 201 02 05 04 ...
Страница 187: ......
Страница 189: ......