5 Operation
Operating Instructions PerioScan
61 29 626 D 3496
D 3496.201.02.05.02
17
båÖäáëÜ
ATTENTION
Flush all water lines to be used for treatment with demin-
eralized water for 20 – 30 s prior to each patient.
ATTENTION
At the beginning of each workday, flush the water lines to
be used for treatment with fresh water – dilution hard-
ness 8° dH (= 1.43 mmol).
This can also be performed by executing the "Continu-
ous flushing" function twice (see page 12). This function
may be activated only with the tip screwed on.
Switching the unit off
If the unit is connected to the fresh water supply system,
you must shut off also the fresh water supply after
switching off the unit with the main switch (1).