Calibration
Operating Instructions Sirona Dental CAD/CAM System
Step 6
Step 6
1. Turn the calibration plate counterclockwise 90° and place it in
position 3.
Note the position of the beveled corner. The grid balls must be seated
as shown in the illustration.
The bottom half of the cylinder is located in the image field.
2. Click the
"Acquire" button.
ª
The scan starts.
ª
The automatic calibration of the XYZ axis alignment is now
completed.
3. Confirm the successful calibration in the inLab software with the
"OK"
button.
4. Remove the calibration plate from the shifting plate.