4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
Sirona Dental Systems GmbH
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
Disinfect the mirror sleeve
1. Press the green plug into the interior of the mirror sleeve in order to
ensure that no particle contamination or fluids penetrate the inside of
the mirror sleeve.
2. Fill the HLD container using one of the following disinfectants for the
high-level disinfection: ProCide-D
®
Plus or CIDEX
®
OPA. Fill the
disinfectant up to the “Omnicam” mark of 50ml.
NOTICE
The plug does not seal the mirror sleeve watertight.
CAUTION
Disinfectants: ProCide-D Plus or CIDEX
®
OPA – Observe the
manufacturers' safety indications!
NOTICE
Only place the mirror sleeve into the container and not the entire
camera.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.