1 Caring for the CEREC Bluecam and prismatic tube
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
båÖäáëÜ
Sterilizing
Do not sterilize the prismatic tube with sapphire glass
The prismatic tube without sapphire glass can be sterilized with hot air
(180°C, 30 min) (not in the autoclave!).
RKI note
Fitting the prismatic tube
Fitting the prismatic tube
1. Pull off the protective cap (E).
2. Carefully refit the prismatic tube until it locks in place.
NOTICE
Not sterilizable!
Do not under any circumstances sterilize the Bluecam or the video
cable!
The camera support cannot be sterilized.
CAUTION
If the Bluecam accidentally falls down, check to make sure that the front
lens and prism are not damaged. If the Bluecam has been damaged, it
must not be used on patients any more.
The Bluecam must be recalibrated in any case.
NOTICE
Risk of damaging the prismatic tube!
The prismatic tube with sapphire glass is not suitable for hot air
sterilization and autoclaving!
The permissible temperature range is: -25°C to +60°C.
NOTICE
For markets where the RKI* guidelines are to be observed
The prismatic tube is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclavable.
*RKI=Robert Koch Institute, Berlin (Germany).
bCL sa / bCB sa
bCL / bCB
NOTICE
Do not use CEREC 2 / CEREC 3 prismatic tubes.
➢ Only use prismatic tubes marked with "bCL sa", "bCB sa", "bCL" or
"bCB".
NOTICE
Risk of damaging the front lens or prism.
➢ The prismatic tube must not touch the front lens.
➢ Push the prismatic tube straight toward the camera body; do not tilt
it.