4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
Sirona Dental Systems GmbH
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
4
Disinfect the mirror sleeve of the CEREC
Omnicam with the HLD set
Fragment: mirror sleeve: no water in camera
The complete process for high-level disinfection (HLD) is as follows:
Cleaning the camera
➢ Cleaning process after using the CEREC Omnicam Camera: Wipe
the camera with the appropriate mirror sleeve so that the superficial
impurities cannot harden and stick to the surface of the mirror sleeve.
Use a lint-free cloth as listed below which has first been soaked in
one of the cleaning products.
Agent approved for USA/not approved for USA
Not approved in the USA
Approved in the USA
Rinsing the exterior of the mirror sleeve
➢ Rinse the exterior of the mirror sleeve.
NOTICE
Make sure that no water gets into the camera when rinsing the mirror
sleeve.
Alpro
● Minuten Spray classic
● Minuten Wipes
● Plasti Sept
● Plasti Sept Wipes
Merz
● Pursept-A
Dürr
● FD 312
Kerr Corporation
● CaviCide
● Cavi Wipes