Figure 10: wiring diagram for the Remote Setpoint U for the CM5
5.8.2 Analogue control signal Remote Current Limit/Current Setpoint
To connect the control signal for the output current to be supplied, a voltage or current source
must be connected between pin 2 and one of the earth pins 15, 16, 17, 18, 19, 24 and/or 25 as
indicated in the diagram below.
Figure 11: wiring diagram for the Remote Setpoint I for the CM5
Note:
In order to be able to use the control input for controlling or limiting the current, the option
4300
must be set to external in the menu (menu item
4320
). If this option is not enabled, the
maximum output current to be supplied will be determined via the menu settings.
5.9 Analogue readout signals
Two analogue signals are available for reading out the high voltage or output current. Both
signals can be configured and read out independently of each other. In order to configure the
outputs the generator menu must be used. For configuring the outputs, see menu items
6610
to
6614
and
6620
to
6624
.
Содержание CM5
Страница 11: ...CM5_UM_ 9752090319_GB_V0_9 11 67 Figure 5 dimensions of the CM5 ...
Страница 13: ...CM5_UM_ 9752090319_GB_V0_9 13 67 Figure 6 CM5 assembly holes ...
Страница 34: ...CM5_UM_ 9752090319_GB_V0_9 34 67 ...
Страница 36: ...CM5_UM_ 9752090319_GB_V0_9 36 67 ...
Страница 39: ...CM5_UM_ 9752090319_GB_V0_9 39 67 ...
Страница 41: ...CM5_UM_ 9752090319_GB_V0_9 41 67 ...
Страница 43: ...CM5_UM_ 9752090319_GB_V0_9 43 67 ...
Страница 45: ...CM5_UM_ 9752090319_GB_V0_9 45 67 ...
Страница 47: ...CM5_UM_ 9752090319_GB_V0_9 47 67 ...
Страница 59: ...CM5_UM_ 9752090319_GB_V0_9 59 67 ...