10
A E R O S T A T S E N T R Y
5 1 0 0 6 0 0 R e v . H
If no instrumentation is available, the unit’s operation can be veri
fi
ed with the
following procedure:
1. Tear off about a 10 inch length of transparent tape. You will note that the tape
is charged and will easily attract to a nearby surface, such as a wall, when
held vertically approximately 1inch from the surface.
2. Pass the tape through the ionized air approximately 1 foot from the ionizer
and again approach the surface with the tape. If it has been neutralized, it will
not attract.
NOTE –
When performing this procedure, do not use a plastic surface such
as acrylic as the surface itself may be charged and attract the tape. A “safe”
surface would be a metal leg or drawer front on a workbench.
Do not try to determine the unit’s operation by the “spark test”. The design
of the balancing circuit makes the “spark Test” inconclusive. Sustained
grounding of the ion emitters may damage the balancing circuit.
Calibration
The Sentry’s ion output is inherently balanced by design. As a result, there are no
calibration adjustments