A E R O S T A T   S E N T R Y

5 1 0 0 6 0 0   R e v .   H

If no instrumentation is available, the unit’s operation can be veri

fi

 ed with the 

following procedure:
1.  Tear off about a 10 inch length of transparent tape.  You will note that the tape 

is charged and will easily attract to a nearby surface, such as a wall, when 
held vertically approximately 1inch from the surface.

2.  Pass the tape through the ionized air approximately 1 foot from the ionizer 

and again approach the surface with the tape.  If it has been neutralized, it will 
not attract.

NOTE – 

When performing this procedure, do not use a plastic surface such 

as acrylic as the surface itself may be charged and attract the tape.  A “safe” 
surface would be a metal leg or drawer front on a workbench.

Do not try to determine the unit’s operation by the “spark test”.  The design 
of the balancing circuit makes the “spark Test” inconclusive.  Sustained 
grounding of the ion emitters may damage the balancing circuit.

Calibration

The Sentry’s ion output is inherently balanced by design.  As a result, there are no 
calibration adjustments

Содержание Aerostat Sentry 4003000

Страница 1: ...AEROSTAT SENTRY 5100600 Rev H Aerostat Sentry Ionizing Blower INSTALLATION AND OPERATING INSTRUCTIONS...

Страница 2: ...AEROSTAT SENTRY 5100600 Rev H TABLE OF CONTENTS 1 DESCRIPTION 3 2 SAFETY 4 3 SPECIFICATIONS 5 4 INSTALLATION 7 5 OPERATION 8 6 MAINTENANCE 9 7 REPLACEMENT PARTS 10 8 WARRANTY 11...

Страница 3: ...ield that injects alternating polarity ions into the airflow The shaped airflow exiting the Sentry carries these ions over a broad area for static elimination The ionizer of the Sentry features a bala...

Страница 4: ...m being returned should be shipped prepaid and packed to provide adequate protection 2 SAFETY NOTE Statements identified with a NOTE indicate precautions necessary to avoid potential equipment failure...

Страница 5: ...elocity Fan Speed 1 ft 2 ft 3 ft 4 ft Low 600 300 180 150 Medium 800 400 220 180 High 1000 500 250 200 Velocity in FPM measured at centerline of air stream Air Flow characteristics 3 x 6 Area Coverage...

Страница 6: ...t discharge time Fan Speed High 1 ft 2 ft 3 ft 4 ft 4 5 6 8 1 ft 1 5 2 5 4 5 5 Center Line 4 5 6 8 1 ft Fan Speed Medium 6 7 9 11 1 ft 1 7 3 5 5 7 Center Line 6 7 9 11 1 ft Fan Speed Low 8 9 10 12 1 f...

Страница 7: ...ration Plug the unit into a standard 2 terminal double grounded receptacle To set the unit up in a bench stand configuration secure the feet to the mounting yoke 1 Place the feet on a table top with t...

Страница 8: ...he Sentry produces an ionized air stream that covers a large area uniformly The time required to neutralize a static charge on an item in this area depends on many factors Two important factors are di...

Страница 9: ...ers 1 Grasp the emitter cleaner between thumb and forefinger 2 Sweep back and forth across the ionized air outlet Recommended frequency of cleaning is once a week The ionized air outlet should remain...

Страница 10: ...and again approach the surface with the tape If it has been neutralized it will not attract NOTE When performing this procedure do not use a plastic surface such as acrylic as the surface itself may b...

Страница 11: ...h 6 4670700 Brush 7 4401510 Ionizer Duct Standoff long 8 4401591 Ionizer Duct Standoff short 9 4103940 Line Cord Assembly 120 V 9 4104393 Line Cord Assembly 230 V 10 4610531 Rocker Switch 11 4610758 3...

Страница 12: ...any goods for any reason contact Simco Ion Customer Service at 215 822 6401 for a Return Authorization Number This number must accompany all returned items This warranty does not apply when the equip...

Отзывы: