1

 

Device description 

The oval gear flow meter of the series VO is a volume displacement flow meter. It is intended 
to be used for the simple and reliable measurement of volumes and flow rate. It has an ex-
tremely robust design and combines years of experience with state-of-the-art technologies.  
The VO can be used in various industries, e.g. mechanical engineering, plant construction, 
food industry, semiconductor industry, environment industry, automotive industry, etc. 
Due to the available material combinations of the housing, oval wheels and gaskets VO is also 
suitable for measuring aggressive or corrosive media. 
Components:  
The VO consists of the two main components, measuring chamber with oval wheels and 
pulse sensors / electronic with display. 

 

  Electronics with display or 

pulse sensors M12x1 (without illustration). 

 

  Upper housing part. 

 

  Process connections (input and output 

side). 

 

  Lower housing part. 

 

  Cable gland (displays 1+2). 

 

  Oval wheels. 

 

  Gasket. 

 
 

 
Versions: 
The VO can be supplied in different types and sizes. Furthermore, the VO can be equipped 
with different display units or a sensor. 
The process connection is available as female 
thread or flange connection. 
The types differ in terms of the combinations of 
materials used in the housing, oval wheels and 
seals. There are 8 sizes for measuring ranges 
from 0.03...700 l/min. 
The display unit is available in three sizes and 
can be mounted on the VO or installed offset on 
the wall. In addition, the VO can also be combined with a sensor or high temperature "sen-
sor".  
Further information on the material combinations, size, display units and sensors can be 
found in our catalogues at "

catalogues.sika.net"

 

 

VO 10 VP with display 3 and  

view from below without lower part 

 

Threaded version  

 

Flanged version 

Содержание VO 015

Страница 1: ...the appliance is resold please provide the operating manual along with it SIKA Ba_VO 06 2015 Oval gear flow meters VO Types VO VA VO VP VO AP Sizes 015 06 1 2 5 10 50 115 Operating manual Translation...

Страница 2: ...ey functions in normal mode 50 6 3 Operating menu 50 6 4 Open collector pulse output displays 2 and 3 56 6 5 Level and parameter operating menu 57 7 Problems 59 8 Maintenance cleaning repairs hazardou...

Страница 3: ...of injury This sign indicates dangers that cause personal injuries that can lead to health defects or cause considerable damage to property CAUTION High temperature This sign indicates dangers result...

Страница 4: ...Electronics with display or pulse sensors M12x1 without illustration Upper housing part Process connections input and output side Lower housing part Cable gland displays 1 2 Oval wheels Gasket Versio...

Страница 5: ...s for completeness Compare the device data with the data on the delivery note and in the order records Device 1x VO according to the order data 2x Process connection protective caps 1x Operating manua...

Страница 6: ...he VO as a safety component The operational safety of the device supplied is only guaranteed by intended use The speci fied limits 10 Technical data may under no circumstances be exceeded Before order...

Страница 7: ...on This can be based on training or relevant tuition The personnel must be aware of this operating manual and have access to it at all times General safety instructions In all work the existing nation...

Страница 8: ...to the measuring chamber wall the input side is separated and sealed from the out put side Between the individual oval wheels and the adjacent measuring chamber wall a movable measuring chamber is fo...

Страница 9: ...ur oval gear flow meters are precision flow meters Inlet and outlet are covered with protective caps against foreign substances Remove caps shortly before putting the device into operation As indicate...

Страница 10: ...ratures Make sure that the installation has cooled down before mounting dismounting Remove any impurities from the pipe work Flush the pipeline when doing so replace the VO with a suitable piece of pi...

Страница 11: ...that are supplied with calibration you receive a test certificate with your device the device specific pulse factor K factor in pulses per litre will be listed This K factor is also specified on the d...

Страница 12: ...ctrical system before connecting the VO 5 2 1 Sensor For connection to evaluation electronics or a PLC or of the user sensor must be powered with 10 30 VDC 3m connection with PVC cable 3 x 0 34 mm 2 L...

Страница 13: ...nd the operating mode A sliding switch A B is provided on the connec tion circuit board for this purpose An external sensor with a cable length of up to 30m can always be connected at terminals 1 and...

Страница 14: ...d to the oval gear flow meter Remove the con nector from the electronics 2 Place the electronics with the display facing downwards on a suitable surface Unscrew the 4 crosshead screws on the underside...

Страница 15: ...tatuses sensor signals pro gramming symbols etc Use the key to select the display Q V1 V2 Q V1 V2 Explanation of the display symbols A or B is displayed for approx 0 5 s when a pulse is generated at t...

Страница 16: ...the key to set the units kg t l or m Once the desired unit has been set press the key to save it Level 6 K Factor The K factor is displayed as 0660 0000 This needs to be changed to 0200 000 The first...

Страница 17: ...w zero is impossible the smallest value that can be displayed at the volume meter is 0 Select operating mode 3 6 3 3 Volume units level 2 The displayed volume unit can be freely selected litters cubic...

Страница 18: ...23 The alpha value is always positive The temperature of the connected PT1000 is recorded and displayed in level 14 The temperature is recorded in 0 5 steps Please observe the permissible temperature...

Страница 19: ...caling Definition of the positions for the following explanations Position 1 far right Position 2 second from right etc Example 1 Setting 1 has been selected in level 24 position 1 Meter reading V1 01...

Страница 20: ...d is preset during production and cannot be changed by the customer The device comes with a printed pro tocol Device Configuration VO Electronics containing all the programmed parameters The password...

Страница 21: ...4 always check and correct the measuring ranges in level 123 and 124 6 3 14 Current simulation level 125 only display 3 Any value between 04 00 and 20 00 mA can be set The set value is output as curr...

Страница 22: ...r voltages of between 5 and max 24 V The cur rent carrying capacity of the output is max 30 mA If a voltage is applied to terminal 5 without resistance the output stage of the electronics will be dest...

Страница 23: ...de 1 A and B at a measuring chamber possibility of return identification 2 A B measurement of the total volume 3 A B measurement of the difference consumption measurement 2 Volume unit Selection of th...

Страница 24: ...perature sensor Factor for density correction 24 Pulse output scaling 0 Original pulses of sensor A pulse duration original pulses 4ms 1 1 pulse per change of the right hand position of V1 right 2 1 p...

Страница 25: ...If the oval gear flow meter will not be in operation for a longer period of time it has to be dismounted thoroughly cleaned and conserved with acid free oil Oval gear flow meters used for liquid food...

Страница 26: ...e if the medium is hazardous to health i e caustic toxic carcinogenic or radioactive etc Please do not return the device if you are not perfectly sure that all media hazardous to health have been clea...

Страница 27: ...steel Aluminium Oval gear Stainless steel PEEK PEEK Bearing Coal PEEK PEEK Axle Stainless steel Stainless steel Stainless steel O ring FKM Standard optional EPDM FEP Process variables Medium to measu...

Страница 28: ...ue Nm Thread Flange 5 5 5 24 35 30 30 55 37 37 60 60 10 1 2 Sensor and display unit Sensor Display 1 Display 2 Display 3 Version Display LCD 8 digit Local on site indication Remote wall mounting Elect...

Страница 29: ...7 245 Bmax D mm 96 101 111 120 125 152 213 271 Installation size mm 73 73 73 90 102 102 184 196 F G G G G G 1 G 1 G 2 G 2 Depending on sensor Process connection flange DIN 1092 1 Form B1 Without displ...

Страница 30: ...Technical data VO 64 SIKA Ba_VO 06 2015 Wall mounting plate...

Страница 31: ...VO Technical data Technical changes reserved 65 10 3Pressure drop...

Страница 32: ...Technical data VO 66 SIKA Ba_VO 06 2015...

Страница 33: ...VO Technical changes reserved 67 For your notes...

Страница 34: ...easuring Instruments Durchflussmesstechnik Flow Measuring Instruments Test und Kalibriertechnik Test and Calibration Instruments SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 D 34260 Kaufungen Germany...

Отзывы: