4.1.6
Frequency measurement and counting
Range
Resolution
Range of measurement Accuracy
Notes
20 kHz
< 0.01 Hz
1 Hz up to 20 kHz
0.005 % of rdg.
•
Temperature coefficient < 5 ppm/ °C from 0 °C to 18 °C and from 28 °C to 50 °C.
•
Triggering level 1V.
•
Scale in beats/min and Hz.
•
Measurement for frequency output and dry contact.
•
In the case of counting, this measurement may be made for a defined time or an infinite
time.
4.1.7
Additional features at measurement mode
Manual or automatic range selection
For the mV, V and Ω functions,
with automatic range selection, the unit selects a higher or
lower range.
Relative measurement
The relative measurement function is used to:
•
programming a reference value other than that of the unit (ZERO function),
•
cancel by measurement or programming a constant or interfering value (TARE function).
Scale correction
The scale correction function performs a conversion between measured electrical quantities
and the physical quantities converted.
Linearisation
Linearisation is used partially to correct errors induced by non linear sensor /c onverter sys-
tems.
Statistics
Display of the minimum, maximum and average value and the number of measurement
points.
The statistics may be reset to zero.