2.2.1
Voltage measurement (DC)
•
Display the
MEASUREMENT CONFIGURATION
dialogue box.
•
Select the
Vdc
measurement function then the correct measurement range using the
function and navigation keys.
•
Confirm with ENTER.
The following ranges are available:
Range
100 mV
1 V
10 V
50 V
Auto
Resolution
1 µV
10 µV
100 µV
1 mV
---
Input impedance
> 10 M
Ω
>
10 MΩ
1 M
Ω
1 M
Ω
---
The DC voltage to be measured is connected between the V and COM terminals.
2.2.2
Current measuring (DC)
•
Display the
MEASUREMENT CONFIGURATION
dialogue box.
•
Select the
Idc
measurement function then the correct measurement range using the
function and navigation keys.
•
Confirm with ENTER.
Depending on the range selected, several modes of measurement are available:
Range
50 mA
4-20 mA
0-20 mA
Resolution
1 µA
1 µA
1 µA
Input impedance
< 30 Ω
< 30
Ω
< 280 Ω
if HART ON
< 30 Ω
< 280 Ω
if HART ON
Loop power supply
no
possible
possible
Set to scale
no
linear or square law
linear or square law
If loop power supply is on, the connection is made between terminals 4-w and mA.