Enter the values and press ENTER.
Use the function keys to continue configuring a sensor:
•
to edit an existing calibration point.
•
to add a calibration point.
•
to delete a calibration point.
Between 1 and 4 calibration points can be entered per sensor.
These calibration points are used to calculate a polynomial c(T) of degree 0 to 3, giving the
sensor's voltage (or resistance) correction at temperature T.
In the specific case where a single calibration point is given, the behavior will differ according
to whether the sensor is a thermocouple or a thermometric resistance:
In the case of a thermocouple, the correction is a fixed voltage deviation.
In the case of a resistive probe, the correction made is an R0 correction.
Sensors declared in this manner are added to the list of couple types (or of probe types) pro-
posed in the measurement function settings dialog box. They appear at the top of the list, in
front of the standard sensors. Their name is preceded by the '*' character, indicating that
these are calibrated sensors.